IM-A-1048H
50
No.
Functions
Initialize
Parameters
Digi
ts
Options/ranges
Initial
values
Remarks
F-07
Smoothing
The number of
moving
average times
10e2
0
0 times
0
None
1
2 times
2 times (0.5msec)
2
4 times
4 times (1msec)
3
8 times
8 times (2msec)
4
16 times
16 times (4msec)
5
32 times
32 times (8msec)
6
64 times
64 times (16msec)
7
128 times
128 times (32msec)
8
256 times
256 times (64msec)
9
512 times
512 times (128msec)
A
1024 times
1024 times (256msec)
b
2048 times
2048 times (512msec)
Minimum
scale
10e1
0
1
0
1 counts.
1
2
2 counts.
2
5
5 counts.
3
10
10 counts.
4
20
20 counts.
5
50
50 counts.
6
100
100 counts.
7
200
200 counts.
8
500
500 counts.
9
1000
1000 counts.
Analog filter
10e0
0
300Hz
0
Cutoff frequency : 300Hz
1
100Hz
Cutoff frequency : 100Hz
2
30Hz
Cutoff frequency : 30Hz
3
10Hz
Cutoff frequency : 10Hz
F-08
Additional
value
-99999 to +99999
0
Display offset
F-09
Automatic
ZERO
compensation
ZERO
compensation
range
0 to 99999
0
The range of the Automatic ZERO
compensation.
Determination
time
0.0 to 9.9
0.0
The determination time of the
Automatic ZERO compensation.
Zero-near-zero
0 to 9
0
The compensation range that
automatically changes the least
significant digit number to zero.
F-10
Hysteresis
0 to 99999
0
Common hysteresis
F-11
BCD output
Logic
For details, see the
"INSTRUCTION MANUAL (FOR
BCD/RS-232C/RS-485/TEDS
OPTIONS)".
F-12
Initialize
Initialize
10e0
0
OFF
0
Not initialized
1
ON
Initialized (excluding the BV and
calibration value)
F-13
BV
selection
Bridge
excitation
voltage
10e0
0
10 V
1
Bridge excitation voltage: 10 V
1
2 V
Bridge excitation voltage : 2 V
F-14
TEDS reading
item
For details, see the
"INSTRUCTION MANUAL (FOR
BCD/RS-232C/RS-485/TEDS
OPTIONS)".