IM-A-1048H
39
Table: FUNCTION SELECTING mode and various functions
FUNCTION
SELECTING
MODE
Key operation and WGA-680A movement
in SETTING mode
Functional description,
supplements
F-01
Original value
The "...." is displayed.
To display the current input value
in mV/V, press the SET key.
A function that displays the
raw output value of the
transducer.
F-02
Actual load
calibration
NOTE
Before executing the calibration function, make sure to
execute the DIGITAL ZERO function with no load applied
to the transducer.
Apply load to the transducer.
Set a desired numeric value
(load value) to be indicated,
and press the SET key.
Range: ±99999 (Initial value: 4000)
Set a decimal point position
and press the SET key.
The "F00000" appears. After enabling or disabling the
key code, the new calibration value appears.
Remove the load.
<To enable the key code>
The "00000" is displayed at the
end.
In this status, set the "00628"
or "11111"and press the SET key.
Then, the calibration value is
updated.
After the settings, the key code is enabled. To update the
calibration value next time, you are required to enter the
key code.
To disable the key code, execute the "F-12".
<To disable the key code>
The "00000" is displayed at the
end.
Press the SET key.
Then, the calibration value is updated.
After the settings, the key code is disabled. You are able to
update the calibration value without entering the key code.
A function that applies the
actual load to the transducer
and sets the indicated value
at that time.
For details, see "4-4-1.
ACTUAL LOAD
CALIBRATION".
The "00628" and "11111" are
key codes. The key codes
prevents the calibration
values from changing due to
malfunctions.
decimal point position