IM-A-1048H
40
FUNCTION
SELECTING
MODE
Key operation and WGA-680A movement
in SETTING mode
Functional description,
supplements
F-03
Sensitivity
registering
calibration
(See below figures)
(1) Rated output
Set the rated output (mV/V)
of the transducer and
press the SET key.
Range: -3.2000 to 3.2000 (Initial value: 2.0000)
(2) Rated capacity
Set a desired numeric value
(rated capacity) to be indicated
and press the SET key.
Range: ±99999 (Initial value: 4000)
(3) Decimal point position
Set a decimal point position
and press the SET key.
Then, the calibration value is updated.
(4) Enable/disable the key code
The "F00000" appears.
After enabling or disabling the key code, the new
calibration value appears.
<To enable the key code>
The "00000" is displayed at the
end.
In this status, set the "00628"
or "11111"and press the SET key.
Then, the calibration value is
updated.
After the settings, the key code is enabled. To update the
calibration value next time, you are required to enter the
key code.
To disable the key code, execute the "F-12".
<To disable the key code>
The "00000" is displayed at the
end.
Press the SET key.
Then, the calibration value is updated.
After the settings, the key code is disabled. You are able to
update the calibration value without entering the key code.
A function that conducts
calibration by registering
the rated output and rated
capacity of the transducer
without applying the actual
load.
For details, see "4-4-2.
Sensitivity
REGISTERING
CALIBRATION".
How to set disable/enable
the key codes is same as
"F-02:
Actual Load Calibration".
The "00628" and "11111"
are key codes. The key
codes prevents the
calibration values from
changing due to
malfunctions.
decimal point position