IM-A-1048H
48
4-12-3. Table of functions
No.
Functions
Initialize
Parameters
Digits
Options/ranges
Initial
values
Remarks
F-01
Original
value
The current input value (mV/V) is
displayed.
F-02
Actual load
calibration
Rated value
-99999 to +99999
4000
Linked to the sensitivity registering
calibration.
Decimal point
position
10e0
None
None
Decimal point: None
10e1
The 10th digit
lights up.
The decimal point is displayed at
the 10th digit.
10e2
The 100th digit
lights up.
The decimal point is displayed at
the 100th digit.
10e3
The 1000th digit
lights up.
The decimal point is displayed at
the 1000th digit.
10e4
The 10000th
digit lights up.
The decimal point is displayed at
the 10000th digit.
F-03
Sensitivity
registering
calibration
Rated output
-3.2000 to +3.2000
2.0000
Rated input (mV/V)
Rated value
-99999 to +99999
4000
The indicated value when the input
is the rated input.
This parameter is linked to the rated
value of the actual load calibration.
Decimal point
position
10e0
None
None
Decimal point: None
10e1
The 10th digit
lights up.
The decimal point is displayed at
the 10th digit.
10e2
The 100th digit
lights up.
The decimal point is displayed at
the 100th digit.
10e3
The 1000th digit
lights up.
The decimal point is displayed at
the 1000th digit.
10e4
The 10000th
digit lights up.
The decimal point is displayed at
the 10000th digit.