IM-A-1048H
49
No.
Functions
Initialize
Parameters
Digits
Options/ranges
Initial
values
Remarks
F-04
Hold mode
Hold mode
10e2
0
Tracking
0
Normal measurements
1
Arbitrary point
hold
Sample hold
2
Peak Hold
(PH)
Peak Hold
3
Interval-specified
PH
Interval-specified peak hold
4
Time-specified
PH
Time-specified peak hold
5
Bottom hold
(BH)
Bottom hold
6
Interval -
specified BH
Interval -specified bottom hold
7
Time-specified
BH
Time-specified bottom hold
8
Peak-to-peak
hold (p-p)
Peak-to-peak hold
9
Interval -
specified P-P
Interval-specified peak-to-peak hold
A
Time-specified
P-P
Time-specified peak-to-peak hold
Display
mode
10e1
0
Normal display
mode
0
Interval-specified and time-
specified only.
1
Hold display
mode
Interval-specified and time-
specified only.
Comparison
mode
10e0
0
Normal
comparison mode
0
Compares data normally.
1
Hold comparison
mode
Compares data during the holding
operations.
Delay time
0.00 to 9.99
0.00
The time to starting the detection
interval from being turned on
HOLD command.
Detect time
0.01 to 9.99
0.01
Time specification PH/BH/P-P Only.
F-05
Self-check
Self-check function
(EEPROM and RAM)
F-06
Analog
output (D/A)
scaling
Automatic
scaling
10e0
0
Manual
1
Automatically not executes the
scaling during the sensitivity
registering calibration.
1
Auto
Automatically executes the scaling
during the sensitivity registering
calibration.
ZERO
indicated
value
-99999 to +99999
0
Indicated value that outputs 0V.
FULL-
SCALE
indicated
value
-99999 to +99999
4000
Indicated value that outputs 10V.