IM-A-1048H
28
a) HOLD modes
(1) Arbitrary point hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A holds the measured value at that moment (arbitrary point) until the
HOLD command is turned OFF.
(2) Peak hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects and displays the peak value until the HOLD command is
turned OFF. The WGA-680A releases the peak value detection and holding
operation, when the hold command is turned OFF or reset command is turned
ON.
(3) Interval-specified peak
hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the peak value until the HOLD command is turned OFF.
The WGA-680A holds and displays the peak value when the HOLD command
is turned OFF.
The WGA-680A releases the holding operation when the RESET command is
turned ON.
(4) Time-specified peak
hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the peak value for the pre-set detect time (detecting
interval). After the lapse of pre-set detecting time, the WGA-680A holds and
displays the peak value. The WGA-680A releases the holding operation when
the RESET command is turned ON.
(5) Bottom hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects and displays the bottom value until the HOLD command is
turned OFF. The WGA-680A releases the bottom value detection and holding
operation, when the hold command is turned OFF or reset command is turned
ON.
(6) Interval-specified
bottom hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the bottom value until the HOLD command is turned OFF.
The WGA-680A holds and displays the bottom value when the HOLD
command is turned OFF.
The WGA-680A releases the holding operation when the RESET command is
turned ON.
(7) Time-specified bottom
hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the bottom value for the pre-set detect time (detecting
interval). After the lapse of pre-set detecting time, the WGA-680A holds and
displays the bottom value. The WGA-680A releases the holding operation when
the RESET command is turned ON.
(8) Peak-to-peak hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects and displays the difference (amplitude), between the peak
value and bottom value, and displays the peak-to-peak value of the difference
until the HOLD command is turned OFF. The WGA-680A releases the peak-to-
peak value detection and holding operation, when the hold command is turned
OFF or reset command is turned ON.
(9) Interval-specified
peak-to-peak hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the peak-to-peak value until the HOLD command is turned
OFF. The WGA-680A holds the peak-to-peak value when the HOLD command
is turned OFF.
The WGA-680A releases the holding operation when the RESET command is
turned ON.
(10) Time-specified
peak-to-peak hold
When the HOLD command is turned ON and after the lapse of delay time, the
WGA-680A detects the peak-to-peak value for the pre-set detect time (detecting
interval). After the lapse of pre-set detecting time, the WGA-680A holds and
displays the peak-to-peak value. The WGA-680A releases the holding operation
when the RESET command is turned ON.