Hinge 1, 2, and 3 Offset
See “How to Enter Calibration Data” on
page 2-42
, for instructions to enter calibration
data.
Each Hinge page includes a Hinge Offset Cali-
bration. Hinge 1 and Hinge 3 are servomotor
driven and are calibrated remotely at the
screens.
Perform the calibration of Hinge 2 first.
Hinge 2 is not driven by a servo motor, but is driven pneumatically. For the Hinge 2
calibration, make mechanical adjustments and enter measurements calibration fields.
X:
+
Offset
Y:
+
Offset
X:
-
Offset
Y:
-
Offset
Hinge 2 location. Non Servo driven
Setup Hinges 1-3 Screens
2-53
990-HQ Operation Manual