7.2.5
Receiver Unit surgiCam® 4K Wireless
Fig. 7-2:
surgiCam® 4K Wireless receiver unit (rear side) with pairing connection (side)
1 HDMI OUT
Video signal output: HDMI
(signals HDMI 2.0 compatible, backward-compatible with HDMI 1.4)
2 25/30 Hz
Sliding switch
camera image frequency
(default: 30 Hz)
3 Pairing
Button for starting the pairing process
4 Pairing, red
Flashing:
Starting pairing
Continuously on: Pairing activated
5 Status, yellow Flashing slowly: Channel search
Flashing rapidly: Establishing connection
Continuously on: Connection established. Source image is shown on the monitor
Lit, with short interruptions: Camera can be installed in the light
6 Power, green Flashing slowly: Receiver unit booting
Flashing rapidly: See “Connection process failed”, page 58
Flashing briefly with longer interruptions: Operating mode reached
7 24 VDC
Power supply connection, 24 VDC, type XP-Power ACM12US24
8 Camera pairing connection
Pairing connection (25-pin built-in female D-Sub connector), at the side
Switching on the receiver unit
•
Energize the receiver unit via the power supply unit (power supply connection (7, Fig. 7-2)).
Pairing of camera and receiver unit
This process is required only in order to form a new pair of camera and receiver unit.
The pairing process takes 2–3 min.
•
Connect the monitor to the receiver unit and ensure that the monitor is switched on.
The wireless connection can be established only if a monitor is connected!