![KLS Martin marLED X Service Manual Download Page 72](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990072.webp)
72
•
Press a flat screwdriver (7) through the gap
against the control panel foil (6)
•
Press out the control panel foil (6), pull it off and
dispose of it.
•
Carefully remove any adhesive residues from the
light housing.
•
Make sure that the keypad attachment surface is
totally free from grease and other residues.
•
Push the flat ribbon cable of the new control
panel foil through the gap (1) into the light casing.
•
Precisely position the control panel foil in the
recess (2) and glue it on.
The contact surfaces of the new control panel foil
are located in the direction of the control
electronics.
•
Open the lock (4).
•
Insert the flat ribbon cable (5) into the
connector (3) in the direction of the arrow.
•
Close lock (4) again.
•
Make sure that the flat ribbon cable is attached
properly and not connected at an angle.