![KLS Martin marLED X Service Manual Download Page 37](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990037.webp)
37
5.3
Addressing of the Light (Control Electronics)
5.3.1
DIP Switches for mLB Address and Terminating Resistor
1 Socket for D-Sub cable
3 DIP switch, 8-fold
5 Firmware version
2 Reset button
4 USB connection
For correct synchronization of multiple lights, the mLB address must be set correctly even in the absence of
a wall control panel.
•
When replacing the control electronics, adopt the DIP switch positions!
Switch, 8-fold (1=ON, 0=OFF): mLB address
Mounting position
DIP 1
DIP 2
Address
MP4/green
1
1
3
MP3/blue
0
1
2
MP2/yellow
1
0
1
MP1/red
0
0
0