![KLS Martin marLED X Service Manual Download Page 117](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990117.webp)
117
b) Checking the safety sleeves/spring plates
AC2000/AC3000
1 Safety sleeve
2 Locking segment
3 Fastening screw M3 x 6 mm for safety sleeve
4 Spring arm socket
5 Redundant locking screw – M4 x 5 mm oval
head screw without washer (acc. to DIN 7985)
OndaSpace
1 Spring arm
3 Label (see below)
4 Spring plate
5 M4 Phillips screws (2 pcs)
2 Locking segment
•
Check for presence of the safety sleeve and its locking or fixing screw.
•
Grease the locking segment (2), see “Final steps”, page 121.
•
Replace missing parts.