5.5
Bluetooth System
Addressing see section 5.3.1 “DIP Switches for mLB Address and Terminating Resistor”, page 37.
Rules for setting the terminating resistor
If an external marTouch® Pro is present in the Bluetooth system, the terminating resistor of the multi-pole
light that communicates with the marTouch® Pro via mLB must be set.
There is no mLB/Bluetooth mixed system!
If a marTouch® Pro is present in a Bluetooth system, it must be connected to the control electronics of a
light via cable.
•
Installation on the light head: Internal cabling in the cardanic system.
•
Installation on wall/table/standard rail (external): Cabling (5–9 poles) via mLB to the master light in the
Bluetooth system.
•
All other lights/subscribers must not be connected to the mLB.