2. Checking potential equalization/protective conductor resistance
Measurement acc. to:
☐
IEC 62353
Others:
N/A
Internal resistance between terminal devices and externally accessible protective
conductor (measuring point with the highest internal resistance) max. 0.3
Ω, see
“Resistance measurement 2”
Ω
☐
If measured value R
imax
>0.3
Ω:
measured value from “Resistance measurement 2”
Ω
minus resistance between central equipotential bonding terminal and externally
accessible protective conductor, see comparative measurement in commissioning
protocol
Ω
Result (max. 0.3
Ω)
Ω
☐
or
Internal resistance between the measuring point on the central bearing shaft and
the externally accessible protective conductor, see “Resistance measuring 3”
Ω
Result (max. 0.3
Ω)
Ω
☐
3. Checking the suspension arm system
OK Not OK N/A
a) Checking and securing the fixing screws
☐
☐
☐
b) Checking the safety sleeves/spring plates
☐
☐
☐
c) Check the locking screws LC spring arm
☐
☐
☐
d) Checking the locking segments
☐
☐
☐
e) Check the locking screws LC spring arm
☐
☐
☐
f) Checking the brake screws
☐
☐
☐
g) Checking the end stops on spring arms and tracking arms
☐
☐
☐
h) Checking the spring arm setting
☐
☐
☐
i) Checking the rating plate and the warning labels
☐
☐
☐
j) Checking the sliding/plug-in contacts in the tracking arm
☐
☐
☐
k) Checking the cover panels
☐
☐
☐
l) Checking the locking ring in case of wall bearing/ceiling tube Ø 70 mm
☐
☐
☐
m) Checking the rolls for damage and proper function in case of mobile lamps
☐
☐
☐
4. Checking the light head
OK Not OK N/A
a) Checking for damage
☐
☐
☐
b) Checking the seals
☐
☐
☐
c) Check the sensoGrip and the control panel
☐
☐
☐
d) Checking the cover glass
☐
☐
☐