4.2.3
Circuit Diagram with Integrated ASPS Changeover Assembly
Power module SMPS 28/6 S1.8 U
100–240 VAC 50–60 Hz
Input current 2–1 A
Output Power max. 180 VA (30 ± 0.75 VDC)
Input N fused (L2) only for IT grids (floating power supply)
Connector
Connector designation
Fuse rating
X1
Input 100–240 VAC, 50–60 Hz
F1/F2
F4
X2
Input Battery 24–36 VDC
T 5 AH, 250 V
T 10 AH, 250 V
X3
Output 30 ± 0.75 VDC
X5
ASPS LED AC (GN)/DC (RD) in the canopy
Terminal connection
X1/X4
X2/X3
Conductor cross-section, flexible leads: max.
2.5 mm
16 mm
Conductor cross-section w. connector sleeve: max.
2.5 mm
16 mm
For
IT
grids (power supply:
I
nsulated, consumer protection: grounding (French:
T
erre)) the input
N fused
is
used. In case of standard grids with grounded neutral conductor, the unprotected N-input is used.