PWX
39
CONFIG Settings
Basic Functions
2
CONFIG parameter details
The CONFIG parameters are explained in detail below.
CF00 Resets the panel settings
Communication
Interface Manual
Sets whether the panel settings will be reset. The following settings will be reset. When the
settings are reset, the PWX750ML settings take on the same values as they do when an
*RST command is received.
You can only specify this parameter when the PWX series is being used as a master unit
.
• Output voltage•
Output on/off at power-on
• Output current•
Constant current control using an
Vext or Rext
(CF05)
• Overvoltage protection•
Constant voltage control using an
Vext or Rext
(CF06)
• Overcurrent protection•
External control setting for turning output on and off (CF09)
• Undervoltage limit•
Voltage and current setting limits (CF15 and CF14)
CF01 Master-slave parallel operation parameter
Sets the total number of units (master unit and slave units) in master-slave parallel operation.
Specify “SLAV” for the units that you want to use as slaves.
This parameter is applied when you restart the PWX750ML.
Changing this setting may affect the settings of the master unit or the slave units that are con-
nected in parallel.
CF02 Power-on status parameter
Sets the condition that the PWX750ML will be in when the power is turned on. This setting is
invalid when you are using an external contact to turn the output off.
This parameter is applied when you restart the PWX750ML.
You can only specify this parameter when the PWX750ML is being used as a master unit
.
See
Value
Parameter description
rSt
Resets the panel settings
See
See
Value
Parameter description
SLAV
Use the PWX as a slave unit.
1 to 4
Set the total number of units (1 to 4; 1 indicates independent operation--factory
default setting).
See
Value
Parameter description
SAFE
The settings that were in use immediately before the POWER switch was turned off
are used.
The output is off (factory default setting).
Auto
The settings that were in use immediately before the POWER switch was turned off
are used.
This includes whether output was on or off.
Forc
The settings that were in use immediately before the POWER switch was turned off
are used.
The output is turned on.
Summary of Contents for PWX750ML
Page 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Page 52: ...52 PWX This page is intentionally blank...
Page 80: ...80 PWX This page is intentionally blank...
Page 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 84: ...84 PWX This page is intentionally blank...
Page 99: ...Appendix A Options B Troubleshooting...
Page 106: ...106 PWX This page is intentionally blank...
Page 109: ......