42
PWX
CONFIG Settings
CF12 Output-on startup state parameter
Sets the operation mode to be prioritized when the output is turned on.
When used as a slave device, CC is automatically prioritized. To use the PWX series as a
standalone unit or master unit after using it as a slave unit, set the operation mode that you
want to use.
If you do not set the operation mode, CC will be prioritized.
You can only specify this parameter when the PWX750ML is being used as a master unit.
CF13 Detection time of OCP activation
Set the time when the overcurrent protection (OCP) is activated after an overcurrent is
detected. Set this in 100 ms steps.
CF14 Current setting limit
Set whether to limit the set output current so that its value does not exceed the set overcur-
rent protection (approximately 95 % of the OCP trip point). When the current setting limit is
set to “oFF,” if you set the current setting limit to “on” when the set OCP is lower than the set
current, the current value will be maintained, and the set OCP will be changed to 105 % of the
current value. The parameter will also be changed if the output is on.
This is set to “on” when the panel settings are reset (CF00) and when a preset memory entry
is recalled.
You can only specify this parameter when the PWX750ML is being used as a master unit.
CF15 Voltage setting limit
Set whether to limit the set output voltage so that its value does not exceed the set overvolt-
age protection (approximately 95 % of the OVP trip point) and is not lower than the set under-
voltage limit. When the voltage setting limit is set to “oFF,” if you set the voltage setting limit to
“on” when the set OVP is lower than the set voltage, the voltage value will be maintained, and
the set OVP will be changed to 105 % of the voltage value. In addition, the set UVL will be
changed to match the voltage value. The parameter will also be changed if the output is on.
This is set to “on” when the panel settings are reset (CF00) and when a preset memory entry
is recalled.
You can only specify this parameter when the PWX750ML is being used as a master unit.
See
Value
Parameter description
CC
CC (constant current) is prioritized.
CV
CV (constant voltage) is prioritized (factory default setting).
See
Value
Parameter description
0
0 ms (factory default setting)
100 to 2000
100 ms to 2000 ms
See
Value
Parameter description
oFF
The setting is not limited.
on
The setting is limited (factory default setting).
See
Value
Parameter description
oFF
The setting is not limited.
The setting is limited (factory default setting).
Summary of Contents for PWX750ML
Page 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Page 52: ...52 PWX This page is intentionally blank...
Page 80: ...80 PWX This page is intentionally blank...
Page 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 84: ...84 PWX This page is intentionally blank...
Page 99: ...Appendix A Options B Troubleshooting...
Page 106: ...106 PWX This page is intentionally blank...
Page 109: ......