PWX
51
2
Basic Functions
Factory Default Settings (Initialization)
If you hold down SHIFT+LOCAL while you turn the POWER switch on, the PWX750ML will
be reset to the factory default settings.
The factory default settings are shown below.
Basic settings
CONFIG settings
Basic item
Parameter description
Output voltage
0 V
Output current
105 % of the rated output current
OVP (overvoltage protection)
112 % of the rated output voltage
OCP (overcurrent protection)
112 % of the rated output current
UVL (undervoltage limit)
0 V
Param.
number
CONFIG parameter
Parameter description
CF01
Master-slave parallel operation parameter
1 (independent operation)
CF02
1
Power-on status parameter
SAFE (panel settings that were in use the
last time the power was turned off)
CF03
Method for clearing OHP, FAN, and AC-FAIL alarms
SAFE (do not clear the alarm automatically)
CF04
1
Memory content display parameter
ON (displayed)
CF05
1
CC control using an Vext or Rext
OFF
CF06
1
CV control using an Vext or Rext
OFF
CF07
1
CV/CC control range parameter
LO (0 V to 5 V)
CF08
Range during voltage and current monitoring
LO (0 V to 5 V)
CF09
1
External control parameter for turning output on and off OFF
CF10
1
External control logic parameter for turning output on
and off
LO (a LOW signal turns output on)
2
CF11
Bleeder on/off parameter
ON (bleeder circuit is on)
CF12
1
Output-on startup state parameter
CV (CV is prioritized)
CF13
Detection time of OCP activation
0 ms
CF14
1
Current setting limit
ON (the setting is limited)
CF15
1
Voltage setting limit
ON (the setting is limited)
CF16
Panel display brightness setting
7
CF17
Communication monitor timer setting
OFF (Communication monitoring timer off)
CF20
Remote interface parameter
LAN
CF21
Command language parameter
SCPI
CF22
Emulation parameter during remote control
NONE (PWX series commands)
CF23
SCPI communication error display parameter
OFF (not displayed)
CF24
Multichannel (VMCB) domain number
0 (the multichannel function is not used)
CF25
Multichannel (VMCB) channel number
0 (the PWX is used as a master unit on the
multichannel network)
CF30
DHCP parameter
ON (a server is used)
CF31
AUTO IP address parameter
ON (the AUTO IP function is used)
CF32
MANUAL IP address parameter
OFF (the MANUAL IP address function is
not used)
CF42
SCPI-RAW protocol TCP port number
5025
CF60
RS232C baud rate parameter
19.2 (kbit/s)
1
You can only specify this parameter when the PWX series is being used as a master unit.
2
0 V to 0.5 V or shorted.
Summary of Contents for PWX750ML
Page 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Page 52: ...52 PWX This page is intentionally blank...
Page 80: ...80 PWX This page is intentionally blank...
Page 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 84: ...84 PWX This page is intentionally blank...
Page 99: ...Appendix A Options B Troubleshooting...
Page 106: ...106 PWX This page is intentionally blank...
Page 109: ......