Midi300V/4 Power Supply 

 

 

Instruction manual 

Midi300V/4 

(FB79530 / 12613546) 

 
 

 

 

 

 

Version 01a   

Dec 2012 

Summary of Contents for FB79530

Page 1: ...1 Midi300V 4 Power Supply Instruction manual Midi300V 4 FB79530 12613546 Version 01a Dec 2012 ...

Page 2: ...2 Packing list Midi300V 4 1x Midi300V 4 Power Supply 1x Power Cord 1x Instruction Manual Signed by Date ...

Page 3: ...e g naked metal wire other than properly insulated power supply cables 3 Do not spill liquid or insert metal objects inside the power supply 4 Never block the ventilation holes or place the unit in any enclosure unless there is adequate ventilation never expose the power supply to a direct heat source 5 Never touch any part of the power supply assembly i e power supply cables or electrophoresis ta...

Page 4: ...nvironmental Conditions This power supply may only be installed and operated only under the following environmental conditions 1 For indoor use only 2 Relative humidity 95 3 Atmospheric pressure 75 kPa 106 kPa 4 Altitude 2000 metres 5 Operating temperature ambient to 40 C 6 Pollution degree 2 7 Mains supply voltage fluctuations up to 10 of the normal voltage Symbols The symbols used on the power s...

Page 5: ...ction 3 Installation Instructions 9 Section 4 Operation Instructions 9 4 1 Control interface 9 4 2 Start the operation 10 4 3 To use the constant current or voltage mode 11 4 4 To adjust the time 11 4 5 Starting and stopping the run 12 4 6 Pausing the run 12 Section 5 Troubleshooting and Maintenance 12 Section 6 Ordering Information 17 Section 7 Warranty 17 ...

Page 6: ...6 Section 1 Introduction 1 1 Overview Front view Rear view Main power socket Fuse holder ON OFF Switch Fan ...

Page 7: ...nals allow simultaneous operation of as many electrophoresis units either at a constant voltage or current setting while the timer function may be set continuously or up to a maximum 999 minutes when an alarm sounds to signify termination of the run A user friendly interface houses a conspicuous 3 digit LED to aid set up as well as a convenient pause resume key a particularly useful feature during...

Page 8: ...c recovery after power failure Yes Safety features No load detection sudden load change detection overload detection ground leak detection over voltage current temperature protection maximum power output detection shrouded plugs and sockets Operating conditions Ambient 40 C 95 humidity Stackable Yes Number of output jacks 4 sets in parallel Regulatory conformity EN 61010 1 CE Dual voltage Yes 100 ...

Page 9: ...ntilated and away from sources of direct heat Section 4 Operation Instructions 4 1 Control interface There are seven buttons and four LED indicators on the faceplate The LED indicates the operation status of the unit a Setup Mode before pressing RUN Start Each illuminated LED light shows the parameter currently selected on the faceplate of the power supply For example if the voltage is selected th...

Page 10: ...rease Select 5 Decrease Select 6 Pause Program 7 Start Stop Program 4 2 To start the operation Note To operate under constant voltage or constant current modes adjust the other parameter to the maximum value For example to operate under constant voltage adjust the current to the maximum output of 700mA before running the power supply with the voltage set at the desired output setting ...

Page 11: ...y inserting the red lead into the red power supply output and the black lead into the black power supply output beneath the front fascia plate 4 3 To use the constant voltage or current mode Press the or buttons to set a constant voltage or current respectively 5 Press the or buttons to adjust the voltage or current in 1V or 1mA increments or decrements to the required output value 6 After adjusti...

Page 12: ... the rear of the power supply 10 The programmed settings will be retained within power supply memory and displayed when the power supply is next switched on 4 6 Pausing the run 11 Press key for temporary access to the gel This interrupts the power supply during the run allowing access to the gel A PAUSE message will flash within the LCD screen while the voltage and current will read at zero value ...

Page 13: ...nsure that it is fits correctly within the power supply Use the correct power cable The fuse has blown Replace the fuse Fuse repeatedly broken Hardware failure Contact Fisher Scientific Operation stops Electrophoresis cables are not connected to the power supply or to the electrophoresis unit s There is a broken circuit in the electrophoresis tank Check the connections to the power supply and the ...

Page 14: ...upply 300V Press the START STOP button to clear the error message Contact Fisher Scientific if the problem persists Err3 Error message power supply stops alarm sounds Ground leak detection 1 Check the electrophoresis system for earth leakage 2 If the Err3 error message persists switch the power on off and restart the application Err6 Error message power supply stops alarm sounds Power supply is ov...

Page 15: ...acing the Fuse For additional fuses please contact Fisher Scientific Ltd To replace the fuse 1 Turn off the main power switch at the rear of power supply and detach the power cord 2 Open the fuse compartment located inside the Power Entry Module by inserting a small flat bladed screwdriver into the slot below the ON OFF switch at the rear of the power supply Gently turn the screwdriver to pry open...

Page 16: ...ap the cover closed Maintenance The MIDI300V 4 Power Supply uses all solid state components and should require no maintenance or recalibration under normal use If the unit must be returned for repair contact your local authorised Fisher Scientific distributor ...

Page 17: ...17 Section 6 Ordering Information Cat No Description FB79530 Fisher Scientific MIDI300V 4 300V 700mA 150W Power Supply 110 240 VAC ...

Page 18: ...d by improper operation Units where repair or modification has been performed by anyone other than Fisher Scientific or an appointed distributor or representative are no longer under warranty from the time the unit was modified Warranty is rendered invalid for those units which contain accessories or repaired parts not supplied by Fisher Scientific or its associated distributors Fisher Scientific ...

Page 19: ...19 NOTES ...

Page 20: ...20 NOTES ...