68
PWX
Pin no.
Signal name
Description
Circuit
1
STATUS COM
This is the status output common.
This is the photocoupler emitter output.
2
CV STATUS
This is set to low level when the PWX750ML is in
constant voltage mode.
This is the photocoupler open collector output.
3
CC STATUS
This is set to low level when the PWX750ML is in
constant current mode.
This is the photocoupler open collector output.
14
ALM STATUS
This is set to low level when a protection function
(OVP/ OVP2/ OCP/ OHP/ FAN/ SEN/ AC-FAIL/
SD) is activated.
This is the photocoupler open collector output.
15
PWR ON STATUS
This is set to low level when the PWX750ML is
turned on (PWR ON STATUS).
This is the photocoupler open collector output.
16
OUT ON STATUS
This is set to low level when the output is turned
on.
This is the photocoupler open collector output.
1
2
3
14
15
16
Summary of Contents for PWX750ML
Page 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Page 52: ...52 PWX This page is intentionally blank...
Page 80: ...80 PWX This page is intentionally blank...
Page 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 84: ...84 PWX This page is intentionally blank...
Page 99: ...Appendix A Options B Troubleshooting...
Page 106: ...106 PWX This page is intentionally blank...
Page 109: ......