PFX2512
89
Preparing to Connect the DUT (Battery)
6
Vo
lt
/
T
h
e
rm
o
m
e
te
r U
n
it
(
O
p
ti
o
n
)
■
Input terminal (connector MC1.5/10-G: Phoenix Contact)
When the PFX2512 is shipped from the factory, connectors are
attached to the input terminals. If they are damaged or lost,
contact your Kikusui agent or distributor.
Using AWG 24 to 20 (0.20 mm
2
to 0.52 mm
2
) cables, construct the following cable assembly.
Temperature measurement (THERMOCOUPLE)
Terminal
number
Signal
Description
4
Cell T Term4+
Thermocouple connection terminal 4+
Cell T Term4-
Thermocouple connection terminal 4-
3
Cell T Term3+
Thermocouple connection terminal 3+
Cell T Term3-
Thermocouple connection terminal 3-
2
Cell T Term2+
Thermocouple connection terminal 2+
Cell T Term2-
Thermocouple connection terminal 2-
1
Cell T Term1+
Thermocouple connection terminal 1+
Cell T Term1-
Thermocouple connection terminal 1-
—
NC
Not connected (reserved)
—
NC
Not connected (reserved)
Voltage measurement (VOLTAGE
SENSING)
Terminal
number
Signal
Description
—
FG
Shield electric potential
4
Cell V Term4+
Cell voltage sensing connection terminal
4+
Cell V Term4-
Cell voltage sensing connection terminal
4-
3
Cell V Term3+
Cell voltage sensing connection terminal
3+
Cell V Term3-
Cell voltage sensing connection terminal
3-
2
Cell V Term2+
Cell voltage sensing connection terminal
2+
Cell V Term2-
Cell voltage sensing connection terminal
2-
1
Cell V Term1+
Cell voltage sensing connection terminal
1+
Cell V Term1-
Cell voltage sensing connection terminal
1-
—
FG
Shield electric potential
84-61-7910
OP02
-
PFX
TERM
+
-
+
-
+
-
+
-
1234
THERMO
COUPLE
+
-
+
-
+
-
+
-
1
FG
234
FG
TERM
MAX20V
V
OLT
A
G
E
SENSING
Insert the wire while you
press this part down with a
flat-blade screwdriver.
Remove the first
7 mm of the
wire’s insulation,
and then insert
the wire here.
Incorrect
Incorrect
Correct
The wire itself is
in contact with the
chassis.
Sensing connector plug
Wire scraps are
in contact with the
chassis.
STRIP-GAUGE
7 mm
Summary of Contents for PFX2500 Series
Page 12: ...12 PFX2512 This page is intentionally blank ...
Page 62: ...62 PFX2512 This page is intentionally blank ...
Page 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Page 96: ...96 PFX2512 This page is intentionally blank ...
Page 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Page 143: ......