PFX2512
139
Troubleshooting
App
.
The system suddenly stopped due to a power outage or some other
cause.
Symptom
Remedy
The PC suddenly stopped or
froze.
Is the PC in standby or sleep mode?
Be sure to disable standby and sleep modes.
— —
Restart the PC, and resume the test from BPChecker3000 Test
Executive. There is no need to reconnect.
For details, see “When a power failure occurs on the host PC side or on
all devices” in “Recovery after Power Failures ” in the BPChecker3000
help file.
Test Executive
—
PFX2512 suddenly stopped.
BPChecker3000 status display shows “IOErr.” Restart the PFX2512 and
reconnect. For details, see “When a power failure occurs on a charge /
discharge unit” in “Recovery after Power Failures” in the
BPChecker3000 help file.
Test Executive
—
The PC and PFX2512 suddenly
stopped.
If both the PC and PFX2512 stop, restart the PFX2512 first and then the
PC.
For details, see “When a power failure occurs on the host PC side or on
all devices” in “Recovery after Power Failures ” in the BPChecker3000
help file.
For details, see the BPChecker3000 help file.
Test Executive
—
After a power outage, recovery
or restarting the test does not
work.
Are you using an uninterruptible power supply for the PC? An
uninterruptible power supply is necessary for closing BPChecker3000
and Windows. If you do not use an uninterruptible power supply,
damage may occur to the PC or Windows.
Also, use the auto shutdown management software on the
uninterruptible power supply to shut down the PC properly.
p. 18
See
?
Help
?
Help
?
Help
Summary of Contents for PFX2500 Series
Page 12: ...12 PFX2512 This page is intentionally blank ...
Page 62: ...62 PFX2512 This page is intentionally blank ...
Page 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Page 96: ...96 PFX2512 This page is intentionally blank ...
Page 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Page 143: ......