32
PFX2512
Connecting the Devices
■
Connecting the chassis terminal
The chassis terminal is a grounding terminal designed for many purposes. It can be effective in
reducing noise and preventing malfunction during large current operation. In the event a cable
or the like makes contact with the chassis, the terminal prevent electric shock accidents. To
perform charge/discharge tests safely, we recommend that you ground the chassis terminal.
For the chassis terminal grounding cable, use a cable that can handle at least the maximum
current that will flow during charge/discharge tests.
■
Connecting the devices
The following shows an example of system configuration which consist of 1set of the Charge/
Discharge System Controller (PFX2512), 1 set the DC power supply (PWR1201ML), and 2sets of
the Electronic load (PLZ1205W) in parallel operation. As for the parallel operation of the
Electronic load units (PLZ1205W), locate the master unit with a position nearest to the PFX2512.
Example of a charge/discharge unit with two electronic loads (PLZ1205Ws) connected in parallel
+
–
PLZ1205W master unit
PLZ1205W slave unit
PWR1201ML
PFX2512
DUT (battery)
DUT cable
(Option: TL08-PFX)
Cable with
solderless
terminal
(included)
Cable with
solderless
terminal
(included)
Parallel operation
signal cable kit
(Option:
PC01-PLZ-5W)
I/F cable for PLZ-5W
Electronic load
(Option:
SC05-PFX)
I/F cable for PWR-01
DC power supply
(Option:
SC07-PFX)
Terminal cover
Example of complete connecting
figure of the rear panel
Summary of Contents for PFX2500 Series
Page 12: ...12 PFX2512 This page is intentionally blank ...
Page 62: ...62 PFX2512 This page is intentionally blank ...
Page 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Page 96: ...96 PFX2512 This page is intentionally blank ...
Page 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Page 143: ......