Recommended Test Equipment
The recommended test equipment for the performance verification and calibration is listed below. If the
exact instrument is not available, substitute calibration standards of equivalent accuracy.
An alternate method is to use the Keysight 3458A 8½-digit Digital Multimeter to measure less accurate,
yet stable, sources. The output value measured from the source can be entered into the instrument as the
target calibration value.
Application
Recommended Equipment
Zero Calibration
Keysight 34172B
DC Voltage
Fluke 5720A
DC Current
Fluke 5720A
Resistance
Fluke 5720A
AC Voltage
Fluke 5720A
AC Current
Fluke 5720A
Frequency
Keysight 33500B Series Function/Arbitrary Waveform Generator
High Current
Fluke 5725A
Capacitance (Optional) SCA-1uF Capacitance Standard
Calibration Adjustment Procedures
526
Keysight True
volt
Series Operating and Service Guide