[SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO {ON|1|OFF|0}
[SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO?
Enables or disables automatic null value selection for AC or DC current measurements.
This parameter setting is not shared between AC and DC measurements. The parameters are independent
for AC and DC measurements.
On the 34461A, 34465A, or 34470A, use the
[SENSe:]CURRent:{AC|DC}:TERMinals
command to spe-
cify which current terminals to use for the measurement.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON.
0 (OFF) or 1 (ON)
Configure AC current measurements, using the null function to subtract 100 mA from the measurements.
Then make two measurements and send them to the instrument's output buffer:
CONF:CURR:AC
CURR:AC:NULL:STAT ON;VAL 100 mA
SAMP:COUN 2
READ?
Typical Response: +1.0400,+1.0400
Make a second set of measurements using automatic null value selection:
CURR:AC:NULL:VAL:AUTO ON
READ?
Typical Response: +0.0000,+0.0100
l
When automatic reference selection is ON, the first measurement made is used as the null value for all
subsequent measurements.
[SENSe:]CURRent:{AC|DC}:NULL:VALue
is set to this value. Automatic
null value selection will be disabled.
l
When automatic null value selection is disabled (OFF), the null value is specified by:
[SENSe:]CURRent:{AC|DC}:NULL:VALue
l
The instrument enables automatic null value selection when the null function is enabled
(
[SENSe:]CURRent:{AC|DC}:NULL:STATe ON
).
l
This parameter is set to its default value after a Factory Reset (
), Instrument Preset
(
), or CONFigure function.
Keysight True
volt
Series Operating and Service Guide
323
SCPI Programming Reference