Procedure A: For Firmware Revision Ending in 02, or Greater
This procedure should be performed by qualified service personnel only.Turn off the
power and remove all measurement leads and other cables, including the power cord,
from the instrument before continuing.
1.
Follow the
to disassemble the instrument.
2.
Locate connector J102 on the front panel board. Short pin A1 to B1 as shown by the orange line
below.
3.
Reassemble the instrument, apply power and turn it on; the instrument will generate an error and
beep (this is normal). Wait 30 to 40 seconds for the instrument to power up completely.
4.
Reset the security code to the factory default:
a.
Press
[Shift]
>
[Utility]
>
Test / Admin
>
Security > New Code
.
b.
Enter AT3446XA as the new passcode.
c.
Press
Done
.
d.
Verify the new code and press
Yes
to confirm the change.
Service and Repair
500
Keysight True
volt
Series Operating and Service Guide