[SENSe:]CAPacitance Subsystem
This subsystem configures capacitance measurements.
Command Summary
[SENSe:]CAPacitance:NULL[:STATe]
[SENSe:]CAPacitance:NULL:VALue
[SENSe:]CAPacitance:NULL:VALue:AUTO
[SENSe:]CAPacitance:RANGe:AUTO
[SENSe:]CAPacitance:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]CAPacitance:NULL[:STATe]?
Turns the null function on or off for capacitance measurements.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON.
0 (OFF) or 1 (ON)
Configure capacitance measurements, turn null function on, use the AUTO null function. Make two meas-
urements and send them to the instrument's output buffer:
CONF:CAP AUTO
CAP:NULL ON
CAP:NULL:VAL:AUTO ON
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l
Enabling the scaling function also enables automatic null value selection (
).
l
To set a fixed null value, use:
[SENSe:]CAPacitance:NULL:VALue
.
l
The instrument disables the null function after a Factory Reset (
), Instrument Preset
(
), or CONFigure function.
SCPI Programming Reference
314
Keysight True
volt
Series Operating and Service Guide