Pin
Location Detail
Low reference: large metal shield
U1006 (large tab on right side as
shown in picture):
1.2 V ± 5%
Low reference: large metal shield
U1004 pin 3:
5 V ± 5%
Low reference: large metal shield
U301:
pin 2: +16.8 V ± 5%
pin 3: -16.4 V ± 5%
Keysight True
volt
Series Operating and Service Guide
481
Service and Repair