ACV Values for 34465A and 34470A DMMs
Input
Error from Nominal
Vrms
Frequency Range
Filter
Quick
Check
24 hours 90 days
1 year
2 years
100 mV
1 kHz
100 mV 200 Hz
± 40 µV
± 60 µV
± 70 µV
± 80 µV
50 kHz
± 80 µV
± 90 µV
± 100 µV ± 110 µV
300 kHz
Q
± 1.1 mV
1 V
1 kHz
1 V
200 Hz
± 400 µV ± 600 µV ± 700 µV ± 800 µV
50 kHz
± 800 µV ± 900 µV
± 1 mV
± 1.1 mV
300 kHz
± 11 mV
0.03 V
1 kHz
10 V
200 Hz
± 2 mV
1 V
1 kHz
200 Hz
± 2.2 mV ± 2.4 mV ± 2.5 mV ± 2.6 mV
10 V
10 Hz
3 Hz
± 4 mV
± 6 mV
± 7 mV
± 8 mV
100 Hz
20 Hz
20 kHz
200 Hz
50 kHz
200 Hz
Q
± 8 mV
± 9 mV
± 10 mV
± 11 mV
100 kHz
± 20 mV
300 kHz
± 110 mV
100 V
1 kHz
100 V
200 Hz
Q
± 40 mV
± 60 mV
± 70 mV
± 80 mV
50 kHz
± 80 mV
± 90 mV ± 100 mV ± 110 mV
70 V
300 kHz
± 800 mV
750 V
1 kHz
750 V
200 Hz
± 750 mV ± 900 mV ± 975 mV
± 1.1 V
210 V
50 kHz
± 330 mV ± 351 mV ± 372 mV ± 393 mV
70 V
300 kHz
± 1.5 V
Q
:
Quick performance verification test point.
ACI Values for 34465A and 34470A DMMs
Input
Error from Nominal
Current, rms Frequency Range
Filter
Quick
Check
24 hours
90 days
1 year 2 years
2 A
1 kHz
3 A
200 Hz
± 5.8 mA
5 kHz
1 A
1 kHz
1 A
200 Hz
± 1.1 mA
± 1.3 mA
± 1.4 mA
5 kHz
100 mA
10 Hz
100 mA
3 Hz
± 110 µA
± 130 µA
± 140 µA
1 kHz
200 Hz
5 kHz
200 Hz
Q
100 µA
1 kHz
10 mA 200 Hz
± 4.1 µA
1 mA
1 kHz
± 4.7 µA
± 4.9 µA
± 5 µA
10 mA
1 kHz
± 11 µA
± 13 µA
± 14 µA
5 kHz
Keysight True
volt
Series Operating and Service Guide
519
Performance Verification