STATus:OPERation[:EVENt]?
Returns the sum of the bits in the event register for the
group. An event
register is a read-only register that latches events from the condition register. While an event bit is set, sub-
sequent events corresponding to that bit are ignored. The register bits are cleared when you read the
register.
Parameter
Typical Return
(none)
+512
Read the event register (bit 9 is set):
STAT:OPER:EVEN?
l
Once a bit is set, it remains set until cleared by reading the event register or by sending
(clear
status).
STATus:PRESet
Clears
enable register and
enable register.
Parameter Typical Return
(none)
(none)
Clear enable register bits:
STAT:PRES
STATus:QUEStionable:CONDition?
Returns the sum of the bits in the condition register for the
group. This register
is read-only; bits are not cleared when read.
A condition register continuously monitors the state of the instrument. Condition register bits are updated
in real time; they are neither latched nor buffered.
Parameter
Typical Return
(none)
+4096
Read the condition register (bit 12 is set):
STAT:QUES:COND?
l
The condition register bits reflect the current condition. If a condition goes away, the corresponding
bit is cleared in the condition register.
SCPI Programming Reference
390
Keysight True
volt
Series Operating and Service Guide