IP Address
You can enter a static IP address for the instrument as a four-byte integer expressed in dot notation. Each
byte is a decimal value, with no leading zeros (for example, 169.254.2.20).
l
If DHCP is on, it attempts to assign an IP address to the instrument. If it fails, Auto-IP attempts to
assign an IP address to the instrument.
l
Contact your LAN administrator to obtain an IP address.
l
This setting is non-volatile; it will not be changed by power cycling or
or
.
l
Front Panel:[Utility] > I/O Config > LAN Settings > Modify Settings
Then set the first softkey to
Manual
and press
IP Address
to enter a new IP address.
l
SCPI:
SYSTem:COMMunicate:LAN:IPADdress "<
l
If you change this parameter, you must either press the
Apply Changes
softkey (front panel) or send
(remote interface) for the change to take effect.
Subnet Mask
Subnetting allows the LAN administrator to subdivide a network to simplify administration and minimize
network traffic. The subnet mask indicates the portion of the host address used to indicate the subnet.
l
Contact your LAN administrator for details.
l
This setting is non-volatile; it will not be changed by power cycling or
or
.
l
Front Panel:[Utility] > I/O Config > LAN Settings > Modify Settings
Then set the first softkey to
Manual
and press
Subnet Mask
to enter a new subnet mask with the
arrow keys (for example: 255.255.0.0).
l
SCPI:
SYSTem:COMMunicate:LAN:SMASk "<
l
If you change this parameter, you must either press the
Apply Changes
softkey (front panel) or send
(remote interface) for the change to take effect.
Keysight True
volt
Series Operating and Service Guide
29
Introduction to the Instrument