CALCulate:AVERage:CLEar[:IMMediate]
Clears all computed statistics: minimum, maximum, average, peak-to-peak, count and standard devi-
ation.
Parameter
Typical Return
(none)
(none)
Clear the stored statistical data:
CALC:AVER:CLE
l
Does not clear measurements in reading memory.
l
Statistics are cleared when the measurement function changes or when any of these commands is
executed:
l
To clear statistics, limits, histogram data, and measurement data, use
.
CALCulate:AVERage[:STATe] {ON|1|OFF|0}
CALCulate:AVERage[:STATe]?
Enables or disables statistics computation.
Parameter
Typical Return
{ON|1|OFF|0}. Default: OFF.
0 (OFF) or 1 (ON)
Return the statistics of 100 frequency measurements:
CONF:FREQ 1.0E3,.001
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
*WAI
CALC:AVER:ALL?
Typical Response: +1.0003,+1.0003,+1.0003,+4.13500000E-01
SCPI Programming Reference
232
Keysight True
volt
Series Operating and Service Guide