Key
Purpose
Configure continuity measurements:
Beeper
: Off or On (default)
Configure diode measurements:
Beeper
: Off or On (default)
Configure 2-wire and 4-wire temperature measurements.
Probe Settings
: RTD 2w, RTD 4w (default), Thermis2w, Thermis4w, TCouple
(34465A/70A only)
Additional settings for probe type RTD 2w or RTD 4w
:
l
R0
: R0 is the nominal resistance of an RTD at 0 °C. Default 100 Ω
l
Low Power
: Disables (Off) or enables (On) low power measurements. Applies only
to the 34465A and 34470A.
Additional settings for probe type Thermis2w and Thermis4w
:
l
Low Power
: Disables (Off) or enables (On) low power measurements. Applies only
to the 34465A and 34470A.
Additional settings for probe type TCouple
:
l
Type
: J (default), K, E, T, N, or R
l
Reference
: Internal or Fixed
l
Offset Adjust
:(Available for an internal reference only). -20 °C to +20 °C. Default:
0 °C.
l
Fixed Offset
: (Available for a fixed reference only)-20 °C to +80 °C. Default: 0 °C.
Aperture NPLC
: 0.02, 0.2, 1, 10, 100. Default: 10 (34460A/61A)
0.02, 0.06, 0.2, 1, 10, 100. Default: 10 (34465A/70A without DIG option)
0.001, 0.002, 0.006, 0.02, 0.06, .2, 1, 10, 100. Default: 10 (34465A/70A with
DIG option)
See
for more information.
Aperture Time
(applies only to the 34465A and 34470A): (
Without
the
DIG option) 200 µs to 1 s (2 µs resolution), default: 100 ms. (
With
the DIG option)
20 µs to 1 s (2 µs resolution), default: 100 ms.
Auto Zero
: Off or On (default) (2-wire measurements only; not available for 4-wire
measurements)
OffstComp
: Off (default) or ON. (RTD 2-wire and RTD 4-wire measurements only)
Open Check
: TCouple measurements only.*
Units
: °C, °F, or K
Start and stop measurements.
Features and Functions
52
Keysight True