SYSTem:COMMunicate:LAN:IPADdress "<
address
>"
SYSTem:COMMunicate:LAN:IPADdress? [{CURRent|STATic}]]
Assigns a static Internet Protocol (IP) address for the instrument. If DHCP is enabled (SYSTem:COMMu-
nicate:LAN:DHCP ON), the specified static IP address is not used. Contact your LAN administrator for
details.
If you change this setting, you must send
to activ-
ate the new setting.
Parameter
Typical Return
Command: "
"
Query: {CURRent|STATic}. Default: CURRent.
"169.254.149.35"
Set a static IP address:
SYST:COMM:LAN:IPAD "169.254.149.35"
SYST:COMM:LAN:UPD
l
sets this parameter to its default value.
l
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
For the query form, specify "CURRent" (default) to read the value currently being used by the instru-
ment. Specify "STATic" to read the value currently stored in nonvolatile memory within the instrument
(may not be the actual address used by the instrument if DHCP is enabled).
l
This setting is non-volatile; it is not changed by power cycling, a Factory Reset (
), or an Instru-
ment Preset (
).
l
This parameter is set to its default value when the instrument is shipped from the factory and after
.
SYSTem:COMMunicate:LAN:MAC?
Returns the instrument's Media Access Control (MAC) address as an ASCII string of 12 hexadecimal char-
acters (0-9 and A-F) enclosed in quotation marks.
Your LAN administrator may need the MAC address to assign a static IP address for this
device.
Parameter
Typical Return
(none)
"0030D3001041"
Return the MAC address:
SYST:COMM:LAN:MAC?
l
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
The MAC address is also known as the link-layer address, the Ethernet (station) address, LANIC ID or
Hardware Address. This is an unchangeable 48-bit address assigned by the manufacturer to each
unique Internet device.
SCPI Programming Reference
412
Keysight True
volt
Series Operating and Service Guide