[SENSe:]CAPacitance:NULL:VALue {<
value
>|MIN|MAX|DEF}
[SENSe:]CAPacitance:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for capacitance measurements. The null value is subtracted from each sample if the
null state is turned on.
Parameter
Typical Return
<
value
> -120 to +120 µF. Default: 0.
-1.20000000E-04
Configure capacitance measurements, turn null function on. Make two measurements and send them to
the instrument's output buffer:
CONF:CAP AUTO
CAP:NULL ON
CAP:NULL:VAL 1.2E-12
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l
Specifying a null value disables automatic null value selection
(
[SENSe:]CAPacitance:NULL:VALue:AUTO OFF
).
l
To use the null value, turn on the null state with the
[Sense:]CAPacitance:NULL[:STATe]
command.
l
This parameter is set to its default value after a Factory Reset (
), Instrument Preset
(
), or CONFigure function.