User’s and Service Guide U3047-90002
43
U3047AE04
Cal Kit Operational Check
Cal Kit Operational Check
The following procedure can be used to confirm 1-Port Cal performance of an ECal Module or
Mechanical Cal Kit. The operation verification limits provided ensure that your U3047AE04 and
PNA-X are operating properly. Refer to
“Verifying a Cal Kit Operational Check Failure” on page
to ensure the RF interface cables are installed and working correctly before returning the
Test Set for repairs.
Verification Limits
Specifications for the Test Set are typical. System performance for the PNA-X and Test Set are
only characteristic and intended as non-warranted information. Only a functional certificate is
provided for the Test Set.
It is recommended that you return your instrument to Keysight Technologies for servicing or
repair if the Test Set and PNA-X performance exceed the operational verification limits.
A periodic calibration is not required. The Operators Check should be performed after System
Setup, or if performance is in question. An N-Port calibration should be performed before
making a measurement. Refer to
If you suspect that your 8-Port configuration is not operating properly,
ensure that all front RF jumper interconnect cables are correctly attached.
Equipment Required
The Keysight U3047AE04 requires that the user be familiar with the equipment listed.
Table 1
3
Equipment List
Description
Qty
N4694A 1.85 mm ECal Module 10 MHz to 70 GHz (Option 00F or M0F)
or
85058B Verification Kit.
1
N5247A Option 400 and 551
1
Set of interconnect cables (PNA-X and Test Set), see
1