User’s and Service Guide U3047-90002
45
U3047AE04
Cal Kit Operational Check
N524xA
1-Port Calibration and Verification Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque to 8
in-lb. For further information refer to the Help menu, see
2. Perform a 1-Port Calibration on Port 1. On the PNA-X, select Response > Cal > Start Cal >
Calibration Wizard.
a. If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b. If using an ECal module, connect the ECal to a PNA-X USB port. Select Use Electronic
Calibration (ECal) > Next.
3. Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal Module
dialog box click Clear All, then select Port 1 > Next > Measure.
4. Ensure the ECal you are using is indicated on the right side of the window.
Figure 34 1-Port Calibration
5. Continue to follow the prompts until the Calibration Completed dialog box appears.
6. At the Calibration Completed prompt, select Save As User Calset, type the name 999.1.
Overwrite the Calset if it already exists and Save.
If you do not have a keyboard, select
Save As User Calset
>
Edit Name
and save as
999.x
.
X is the port number you are calibrating. See
Use the numeric keypad on the PNA-X front panel to enter “999.1.”
Figure 35 Calibration Complete