Controlling the U3047AE04 with N5247A
If the PNA-X has Option 029 or H29 (noise figure capability) verify that the
path configuration is set to Default. If the PNA-X is not in Default
configuration, ports 5 and 9 will not operate correctly. Select
Trace/Chan >
Channel
>
Hardware Setup > Path Cong
... and in the drop-down menu select
Default
>
OK
. Save this configuration as a User Preset by selecting
Save
>
User Preset
>
Save current state as User Preset
. Do not use the factory
Preset (User Preset Off), the PNA-X will return to Option 029 path
configuration.
Figure 24 RF Path Configuration
1. On the PNA-X select Response > Cal Wizard.
a. If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b. If using an ECal module, connect the ECal to an available PNA-X USB port and select
Use Electronic Calibration (ECal) > Next.
Figure 25 Begin Calibration