User’s and Service Guide U3047-90002
21
U3047AE04
Controlling the U3047AE04 with N5247A
PNA-X Multiport Mode
The PNA-X requires Option 551 for multiport mode. The PNA-X must have the Test Set file
installed to enable the PNA-X to control the Test Set. The PNA-X multiport mode allows you
to select measurement ports, complete an N-Port calibration, operate control lines, and
rename test ports. Refer to the Help menu for more information.
Multiport Mode Operation
1. The Option 551 must be installed for multiport capability. To access the multiport application
select Utility > System > Configure > Multiport Capability.
2. Select U3070AE04_PNAX_p4 (8-Port System) from the drop-down menu and select Restart
as a multiport PNA with this testset > OK. The PNA-X will restart the network application with
the Test Set interface features.
If the Test Set is not available in the drop-down list, it will be necessary for you to copy the
required Test Set file to the PNA-X hard drive. The current version of the Test Set files are
available on the web at
http://na.support.keysight.com/multiport
Copy the appropriate file to c:\program files\Agilent\Network Analyzer\testsets directory.
Figure 17 Test Set Selection