User’s and Service Guide U3047-90002
35
U3047AE04
Address and Data Values
Address and
Data
Values
Setting the Test Port Path
s
with Address and Data
below
for information to set the internal switch paths
of the Test Set.
The address is the first value in the Test Set I/O control or GPIB data command. The second
value
set
s the source and receiver paths of the ports. To select a Test Set port configuration
both switches must be set to complete the source and receiver paths. Refer to
“GPIB Control Mode” on page 32
If the ports have different addresses, two separate address data commands must be used.
Port 5 is the Source and Port 6 is the Receiver.
Source Port 5 = address 0, data 1 and Receiver Port 6 = address 16, data 16.
Two separate commands must to be sent, you may use the same dialog box. The entry will be
0.1 in one command line, and 16.16 in the second line.
Example 2:
If the ports have the same address, only one command is needed.
Port 1 is the Source and Ports 5 is the Receiver.
Source Port 1 = address 0, data 0 and Receiver Port 5 = address 0, data 16. The data values are
added together, the entry will be 0.17.
Figure 40, “U3047AE04 RF Block Diagram,”
for in-depth RF path information.
Table
9
Test Port Address and Data Values
Address
Source
Path
Receiver
Path
Data
0
1
0
16
0
Ports
1
5
1
5
16
Ports
3
6
3
6
32
Ports
4
7
4
7
64
Ports
2
8
2
8