User’s and Service Guide U3047-90002
41
Control Lines
U3047AE04
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the PNA is in its
Standalone Mode.
Setting the PNA to Standalone Mode
1. Select Utility > System > Configure > Multiport Capability. In the dialog box select
Restart as a standalone PNA > OK.
As described in
“Setting the Test Port Paths with Address and Data” on page 35
<address>.<data> values are determined in a similar manor, with the following exceptions:
•
will be used.
•
The <address> value is always = 112
Table 1
2
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8