User’s and Service Guide U3047-90002
15
U3047AE04
System Setup
RF Interface Cable Connections
(U3021-60101)
illustrates the cable configuration of the Test Set to the PNA-X. The
cables have been supplied with Kit U3021-60101.
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8) on the
PNA-X. The RCVR R1 to SOURCE OUT reference loop jumpers (x4) remain on the front
panel.
2.
Connect the RF interconnect cables (U3047-20016 and U3047-20017) from the PNA-X to
the Test Set in the order listed. As you are connecting each cable, torque to 8 in-lb. Refer to
Over-torque
will cause damage to the Test Set and may cause
connectors to spin or become loose.
illustrates the final two digits of the part number for each cable. The cables must be
connected in the numeric order listed in
Figure 10
PNA-X RF Interface Cable Connections
Table
7
N5247A RF Interface Cable Connection
Numeric
Order
RF Cables
From PNA-X
To Test Set
1
U3047-20016
CPLR ARM
CPLR ARM
2
U3047-20016
RCVR IN
RCVR OUT
3
U3047-20017
CPLR THRU
CPLR THRU
4
U3047-20017
SOURCE OUT
SOURCE IN
16 (x8)
17 (x8)