Calibrating
6
N7020A and N7024A Power Rail Probes User’s Guide
59
N7020A / N7024A Probe’s DC Attenuation/Offset Calibration Using
N7023A Probe Browser
1
Turn on the oscilloscope and connect the N7023A power rail probe browser with
the N7020A / N7024A probe attached to one of the oscilloscope’s input channels.
2
Allow the oscilloscope to warm up for 20 minutes.
3
If the oscilloscope needs calibration, perform a user calibration before the probe
calibration. On the oscilloscope, click
Utilities
>
Calibration
.
4
Connect the ground lugs and the output of the N7023A power rail probe browser
to the oscilloscope’s front-panel PROBE COMP connector.
5
On the oscilloscope, click
Setup
>
Probes
.
6
In the Probe Calibration dialog box, select the tab representing the channel that
has the probe attached.
7
In the dialog box, select the type of calibration. Click
Start
and follow the
instructions shown on the oscilloscope.
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...