82
N7020A and N7024A Power Rail Probes User’s Guide
8
Performance Verification
4
Connect the other end of the calibration coax cable to the Cal Out connector on
the front panel of the oscilloscope.
Figure 28
Calibration setup with the calibration cable
5
Press
[Auto Scale]
on the front panel.
6
Choose
Setup > Acquisition
, then select the
Enabled
checkbox for
Averaging
and set
the
# of averages
to
1024
.
7
Set timebase to
2ns/div
.
8
Choose
Setup
>
Math Functions...
or
Math
>
Functions...
.
9
In the
Function
dialog box, select the tab of the function
f1
to define it.
10
From the
Function 1
listbox, select
Math
and then
Differentiate
.
NOTE
Use the 3.5mm or 2.92mm SMA female to female adapters to make
these connections to Cal Out and input channel of oscilloscope.
Calibration
Cable
Cal Out
Connector
SMA f-f
Adapter
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...