Performance Verification
8
N7020A and N7024A Power Rail Probes User’s Guide
85
19
Check the frequency response of the N7024A probe. To do this, perform the
following steps.
a
Disconnect the calibration coax cable and any adapters that you attached to
the input channel of the oscilloscope.
b
Connect the N7024A probe to the input channel and calibration coax cable.
You may need to use the N5442A adapter with the probe in case you are
using a V-series or a Z-series oscilloscope.
Figure 31
Calibration setup with the N7024A probe and calibration cable
c
Choose
Setup
>
Probe Configuration...
.
d
Select the tab of the oscilloscope’s input channel to which the probe is
connected.
e
In the
Probe Head
block, click
Select Head...
. Select
Generic:User Cable
to
configure the oscilloscope to not consider any loss from the N7022A main
cable or any probe browsers. Click
OK
and close the dialog.
Calibration
Cable
SMA f-f
Adapter
N7024A
Probe
N5422A
Adapter
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...