86
N7020A and N7024A Power Rail Probes User’s Guide
8
Performance Verification
f
Set the channel offset for the oscilloscope to
-640mV
and the trigger level to
-640mV
as displayed in
Figure 32
.
g
The f2 trace now has the frequency response with the N7024A probe
attached. The already saved and displayed memory 1 trace has the
frequency response of the coax cable. For comparison, grab the f2 trace and
align it with the memory 1 trace, if needed as displayed in
Figure 32
.
h
The frequency response of the N7014A probe is the difference between the
f2 trace and memory 1 trace. Verify whether or not this difference is less
than
3db
at 6 GHz or below. If it is less, then the probe passes the test.
Notice that in the figure below, the N7024A probe passed the bandwidth
verification test as the frequency response of the probe s not above 3db at 6
GHz or below.
Figure 32
Frequency response of the N7024A probe
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...