84
N7020A and N7024A Power Rail Probes User’s Guide
8
Performance Verification
18
Record and save the frequency response of the Cal Out step through the attached
calibration coax cable. To do this, perform the following steps.
a
The f2 trace is displayed at the bottom of the screen. Drag the f2 trace to the
center graticule.
b
Save the frequency response of the coax cable by right-clicking the f2 trace
and saving it to
memory 1
. This will allow you to later compare the saved
frequency response of the Cal Out step through the attached calibration
coax cable with the frequency response of the N7024A probe (done in steps
that follow).
c
For ease of comparison, drag the memory 1 trace into the same window as
the f2 trace. Close the window that was initially displayed for memory 1.
Figure 30
Frequency response of the Cal Out step through the attached coax cable
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...