Performance Verification
8
N7020A and N7024A Power Rail Probes User’s Guide
89
8
Consider the N7022A main cable as damaged if either of the following is true:
•
The difference between the f2 trace and memory 1 trace is more than
3db
at
6 GHz or below.
•
The difference between the f2 trace and memory 1 trace is less than
3db
at 6
GHz or below but the f2 trace is clearly different from the f2 trace displayed
in the sample figure below.
Figure 34
Frequency response of the N7024A probe with the N7022A main cable
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...