![Keysight Technologies N7020A User Manual Download Page 70](http://html.mh-extra.com/html/keysight-technologies/n7020a/n7020a_user-manual_1962049070.webp)
70
N7020A and N7024A Power Rail Probes User’s Guide
7
Performance Plots
N7024A Input Impedance Plots
This section provides the input impedance plots for various probe configurations
available for N7024A based on the different probe and probe head combinations.
N7024A Simplified Input Impedance Model
Figure 21
shows the model used in the measured and modeled input impedance
plots provided for the N7024A probe in this section.
In this model, the value of the inductance L varies for various supported probe
configurations. This value is provided with the input impedance plot of each of
these probe configurations in this section.
Figure 21
Input Impedance Model for the N7024A Probe
Input Impedance Plot for N7022A Main Cable and N7024A Probe
Combination
This combination supports the highest bandwidth configuration for the N7024A
probe. In this combination, the value of the inductance L is zero.
Figure 22
shows the input impedance looking into the N7022A cable. Any added
transmission line and/or connection parasitic on the DUT will modify this
impedance.
Summary of Contents for N7020A
Page 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Page 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Page 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Page 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Page 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Page 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Page 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Page 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...