Full performance verification for 90000 A-Series oscilloscopes consists of three
main procedures:
1
Performing the internal oscilloscope self tests to ensure the measurement
system is functioning properly. To perform the self tests, Click
Utilities > Self
Test...
. Then select
Scope SelfTest
from the
Available Self Test
drop-down list box,
click
Start
, and follow the instructions on the screen. If any of the self tests fail,
ensure that the failure is diagnosed and repaired before calibrating and testing
performance.
2
Calibrating the oscilloscope, as described in chapter 2.
3
Testing the oscilloscope to ensure that it is performing to specification. This
chapter describes the performance test procedures.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......