56
90000 A-Series Service Guide
4
Troubleshooting
It is also possible that there is a short circuit on the bulk +12 V supply trace
somewhere in the system, which would also cause the oscilloscope to not power
on. If so, you need to isolate the assembly that is shorted. You do not need to
remove the acquisition or backplane assemblies to perform the following steps.
1
With the power turned off, disconnect the bulk +12 V supply, but do not remove
the supply from the oscilloscope.
Power supply disconnect
2
On the backplane assembly, measure the resistance across the FET (Q1409,
located in the upper right corner of the backplane) b12 V and ground
as shown below, or measure the resistance across the bulk supply input pins on
the backplane assembly. Make sure the red lead is plugged into the HI input
and the black lead is plugged into the LO or COMMON input of the DVM. On a
good backplane assembly, you should measure about 700 ohms. If there is a
short on this board, you would measure something very near zero ohms.
Measuring the backplane resistance b12 V and ground across Q1409
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......